C-Reactive
Protein (CRP): The Differences Amongst Various Assays
(posted March 15, 2006)
(Back to Safety Tips for Laboratorians)
It has been brought to the Food and Drug Administration's attention
that there may be confusion in the laboratory community, from clinicians
and from manufacturers on the indications for use of different
C-Reactive Protein (CRP) assays. This communication is intended
to provide information about the differences of conventional CRP,
high sensitivity CRP (hsCRP) and cardiac CRP (cCRP) assays and
the regulatory need to designate a CRP assay specifically for cardiovascular
risk indications for use. Much of this information is provided
in Guidance for Industry and FDA Staff - Review Criteria for
Assessment of C Reactive Protein (CRP), High Sensitivity C-Reactive
Protein (hsCRP) and Cardiac C-Reactive Protein (cCRP) Assays which
can be found at the following website: http://www.fda.gov/cdrh/oivd/guidance/1246.html.
In brief, the difference between hsCRP and cCRP is not in the
analyte itself, but in the analytical performance which has been
documented through submitted evidence. FDA is responsible for approving
the device labeling (including the specifications stated in the
package insert). When the device specifications indicate that the
device can be used for multiple clinical purposes (e.g., to diagnose
two distinct conditions or diseases), FDA requires appropriate
evidence to support each clinical purpose. When, as in the case
of CRP which is submitted with evidence of efficacy for cardiovascular
risk stratification, FDA could have chosen to approve “High
Sensitivity CRP-with-evidence-of-efficacy-for-identification-and-assessment/stratification-of-individuals-at-risk-for-future-cardiovascular-disease,” but
it chose instead to use the shorthand “Cardiac CRP (cCRP)”to
distinguish such assays from “High Sensitivity CRP (hsCRP)” for
which there is no evidence of efficacy in cardiovascular risk assessment/stratification.
No hsCRP assay has been allowed to have this indication (cCRP)
without supporting clinical studies or bridging studies to a device
which was the subject of the clinical studies. An hsCRP assay will
need to submit a premarket notification and follow the guidance
in order to expand their indications for use statement to include for
use as an aid in the identification and stratification (assessment)
of individuals at risk for future cardiovascular disease. When
used in conjunction with traditional clinical laboratory evaluation
of acute coronary syndromes, cCRP may be useful as an independent
marker of prognosis for recurrent events in patients with stable
coronary disease or acute coronary syndrome
The designation “cCRP” is intended solely to identify
IVDs which meet the performance specifications as described above,
and use of this name is not a requirement imposed on laboratories
performing this test. Since the name “hsCRP” has been
in use when the test is used for cardiovascular risk assessment/stratification,
FDA understands that laboratories may choose to continue to use
this name. However, if such testing is performed using a non-cCRP
kit, such use would constitute an off-label, or unapproved, use
of the device. While physicians may order and use tests based on
their experience and knowledge, regardless of FDA approval, the
manufacturer may not promote or encourage such off label use. Further,
under the provisions of CLIA, the laboratory must validate the
test performance if it performs the test in a manner not consistent
with the manufacturer’s recommendations and intended use.
|